RT612-ST Automated Dissolution Test System
The Raytor RT612-ST represents the pinnacle of high-throughput pharmaceutical analysis, offering a fully integrated 12 position automatic sampling dissolution test system. Designed for laboratories in Malaysia, or across Southeast Asia that require maximum efficiency, this system automates complex sampling workflows while maintaining the highest levels of precision and regulatory compliance.
By combining the 12-position capacity of the RT6 series with advanced robotics, the RT612-ST eliminates manual sampling variability and significantly increases laboratory productivity.
By combining the 12-position capacity of the RT6 series with advanced robotics, the RT612-ST eliminates manual sampling variability and significantly increases laboratory productivity.
Key Automation & Sampling Features
The RT612-ST is built to handle high-volume testing with unmatched accuracy:
- High-Precision Syringe Pumps: Utilizes 12-channel high-accuracy syringe pumps to ensure sampling volume error is kept below ±1%, providing superior repeatability over manual methods.
- Intelligent Fluidics: Features a pipeline automatic cleaning and emptying function, ensuring the system is purged and ready between time points to prevent cross-contamination.
- Automated Needle Positioning: The system automatically lowers sampling needles and pre-filter heads to the exact pharmacopoeial position at every specified time point, ensuring compliance with USP and ChP guidelines.
- Dual-Stage On-line Filtration: Supports secondary on-line filtration with customizable primary filtrate volume settings.
- Massive Collection Capacity: Includes a sample collection rack with at least 120 positions, allowing for extended unattended testing periods.
- Direct Analytical Integration: Supports automatic sampling and direct injection into 1.5mL HPLC vials, streamlining the transition from dissolution to analysis.
Ready to automate your high-throughput dissolution workflow? Contact LMS Scientific today for a technical consultation.
REQUEST A QUOTE / ENQUIRY FORM
SPECIFICATIONS
VIDEO
SPECIFICATIONS
- Number of Positions: 12 Dissolving Cup Positions
- Speed Range: 0-300 rpm (RPM)
- Speed Resolution: 0.01 rpm
- Steady Speed Error: ≤±0.3RPM
- Temperature Range: Room Temperature to 50°C
- Temperature Resolution: 0.01°C
- Temperature Accuracy: ≤±0.20°C
- Vessel Verticality: 90°±0.5°
- Shaft Verticality: 90°±0.5°
- Centering Deviation: < ±2.0 mm
- Depth Positioning Deviation: < ±1.0 mm
- Shaft Wobble: < ±1.0 mm
- Basket Wobble: < ±1.0 mm
- Software: Supports FDA 21 CFR Part 11 requirements
- Maximum Settable Sample Number: 20
- Sampling Range: 1 ~ 20ml
- Minimum Sampling Interval: First Sampling Point: 3 minutes, Follow up Routine Sampling Point: 5 minutes
- Maximum Working Time: 720 hours
- Sampling Precision: ≤±1%
VIDEO
Master the operation of the Raytor RT612-ST Automatic Dissolution System with this comprehensive, step-by-step tutorial. This video is designed for laboratory professionals and researchers who require precision and automation in their drug release studies.
The RT612-ST is a high-performance 12-position automatic sampling dissolution test system that streamlines the testing workflow, reduces manual error, and ensures full compliance with USP, EP, and ChP standards.
The RT612-ST is a high-performance 12-position automatic sampling dissolution test system that streamlines the testing workflow, reduces manual error, and ensures full compliance with USP, EP, and ChP standards.
DISSOLUTION TESTING
DISSOLUTION TESTING
Menu