RT614-ST Automated Dissolution Test System
High-Throughput Pharmaceutical Testing: Raytor RT614-ST
The Raytor RT614-ST is a premier 14-position automated dissolution system specifically engineered for high-capacity pharmaceutical laboratories in Malaysia. By expanding the testing platform to 14 positions, this system allows for simultaneous analysis of larger batches, standards, and blanks, significantly increasing laboratory throughput without compromising on precision.
This automated system fully complies with international regulatory standards, including the USP, EP, ChP, and FDA DPA-LOP.002 guidelines.
This automated system fully complies with international regulatory standards, including the USP, EP, ChP, and FDA DPA-LOP.002 guidelines.
Advanced Automation & High-Capacity Sampling
The RT614-ST is designed to handle complex drug release studies with minimal manual intervention:
- Precision Syringe Pump: Delivers high-accuracy sampling with a volume error of less than ±1%, ensuring reproducible data for critical formulations.
- 210-Position Sample Rack: Features a high-capacity collection system that allows for extensive, unattended multi-time-point sampling.
- Automated Fluid Management: Integrated pipeline automatic cleaning and automatic emptying functions prevent cross-contamination and simplify maintenance between runs.
- Dual-Stage Filtration: Supports secondary on-line filtration with a minimum level of 0.45µm, with customizable primary filtrate volumes adjustable within each dissolution method.
- Direct Injection Capability: Supports optional automatic sampling and direct injection into pre-opened liquid phase vials for immediate analytical transition.
- Intelligent Positioning: Automatically lowers sampling needles and pre-filter heads to the exact pharmacopoeial position at programmed time points.
Data Integrity & Compliance
The RT614-ST is built for a digital-first regulatory environment:
- Audit Trail: Software includes comprehensive audit trail functions and generates electronic experiment reports in PDF format.
- User Management: Supports at least 300 preset accounts with three customizable levels of hierarchical authority control.
- Method Storage: Stores at least 300 experimental methods for rapid recall and consistent testing parameters.
Local Support by LMS Scientific Solution Sdn Bhd
As your partner for Raytor Instruments in the region, LMS Scientific Solution Sdn Bhd provides end-to-end service for your RT614-ST:
- Installation & Validation: Full IQ/OQ/PQ service and documentation included to ensure immediate compliance.
- Maintenance: On-site training for use, cleaning, and maintenance, backed by a dedicated after-sales service line.
- Warranty: 1-year comprehensive quality assurance period for non-human factors.
REQUEST A QUOTE / ENQUIRY FORM
SPECIFICATIONS
VIDEO
SPECIFICATIONS
- Number of Positions: 14 Dissolving Cup Positions
- Speed Range: 0-300 rpm (RPM)
- Speed Resolution: 0.01 rpm
- Steady Speed Error: ≤±0.3RPM
- Temperature Range: Room Temperature to 50°C
- Temperature Resolution: 0.01°C
- Temperature Accuracy: ≤±0.20°C
- Vessel Verticality: 90°±0.5°
- Shaft Verticality: 90°±0.5°
- Centering Deviation: < ±2.0 mm
- Depth Positioning Deviation: < ±1.0 mm
- Shaft Wobble: < ±0.3 mm
- Basket Wobble: < ±0.3 mm
- Software: Supports FDA 21 CFR Part 11 requirements
- Number of Sampling lines: 14
- Number of test tube rack holes: 15 x 14
- Maximum Settable Sample Number: 36
- Sampling Range: 1 ~ 20ml
- Minimum Sampling Interval: First Sampling Point: 3 minutes, Follow up Routine Sampling Point: 5 minutes
- Maximum Working Time: 720 hours
- Sampling Precision: ≤±1%
VIDEO
This video provides a comprehensive step-by-step guide to the operation of the Raytor RT614-ST Automated Dissolution System, demonstrating its efficiency in pharmaceutical testing. It covers the complete workflow for both paddle and basket methods, starting from system login and water bath preparation to automated sampling and post-test maintenance.
Key features demonstrated in the video include:
Key features demonstrated in the video include:
- Method Configuration: Selection and management of up to 300 experimental methods via the digital interface.
- Automated Dosing & Immersion: Synchronized sample delivery for paddle experiments and automatic platform descent for basket immersion.
- Precision Sampling: Hands-free sample collection with automated needle positioning at exact pharmacopoeial heights.
- Integrated Cleaning: Utilization of the “Auto Cleaning” function to ensure thorough pipeline decontamination between runs.