RT600-ST Automated Dissolution System
The Raytor RT600-ST is a state-of-the-art 8 position automatic sampling dissolution test system designed to automate the most labor-intensive aspects of pharmaceutical testing. By integrating advanced robotics with high-precision fluidics, this system ensures total compliance with USP, EP, and ChP standards while significantly reducing human error.
The RT600-ST is the ideal solution for high-throughput laboratories requiring repeatable, unattended dissolution testing for both paddle and basket methods.
The RT600-ST is the ideal solution for high-throughput laboratories requiring repeatable, unattended dissolution testing for both paddle and basket methods.
Advanced Automation & Sampling Features
The RT600-ST system is engineered for superior accuracy and hands-free operation:
- Precision Syringe Pump Sampling: Utilizes a high-accuracy syringe pump to ensure sampling errors are kept below ±1%.
- Intelligent Fluid Management: Features a pipeline automatic cleaning and automatic emptying function to prevent cross-contamination between time points.
- Dual-Stage Filtration: Supports secondary on-line filtration, allowing the volume of primary filtrate to be customized within the dissolution method.
- High-Capacity Collection: Equipped with a sample collection rack providing at least 120 positions for extended testing runs.
- Direct Injection Capability: Supports automatic sampling and direct injection into liquid phase vials, with options for pre-opened vials.
- Automated Needle Positioning: The system automatically lowers sampling needles and pre-filter heads to the exact pharmacopoeia position at every specified time point
REQUEST A QUOTE / ENQUIRY FORM
SPECIFICATIONS
VIDEO
SPECIFICATIONS
- Number of Positions: 8 Dissolving Cup Positions
- Speed Range: 0-300 rpm (RPM)
- Speed Resolution: 0.01 rpm
- Steady Speed Error: ≤±0.3RPM
- Temperature Range: Room Temperature to 50°C
- Temperature Resolution: 0.01°C
- Temperature Accuracy: ≤±0.20°C
- Vessel Verticality: 90°±0.5°
- Shaft Verticality: 90°±0.5°
- Centering Deviation: < ±2.0 mm
- Depth Positioning Deviation: < ±1.0 mm
- Shaft Wobble: < ±1.0 mm
- Basket Wobble: < ±1.0 mm
- Software: Supports FDA 21 CFR Part 11 requirements
- Maximum Settable Sample Number: 20
- Sampling Range: 1 ~ 20ml
- Minimum Sampling Interval: First Sampling Point: 3 minutes, Follow up Routine Sampling Point: 5 minutes
- Maximum Working Time: 720 hours
- Sampling Precision: ≤±1%
VIDEO
Discover the power of precision with the Raytor RT6 General Dissolution System. This video showcases our automated dissolution apparatus, fully compliant with international standards like USP and ChP. Watch to see how we streamline drug release testing with high-precision syringe pumps and automated sampling workflows.
DISSOLUTION TESTING
DISSOLUTION TESTING
Menu